My research developed measurement methods for ceramic processing, and applied them to additively manufactured ceramics. The work spaned method development, custom instrumentation, and characterization, and it was supported by the Office of Naval Research and by the Air Force Research Laboratory through America Makes.
Sintering begins before it can be seen. Interparticle necks form and grow long before the density change is measurable by conventional means, and the properties of the finished part are decided in that window. My dissertation developed a non-contact laser-ultrasonic method to measure it.
I designed the experimental platform and the sample-alignment hardware around a custom laser-ultrasonic system with pulsed-laser generation and fiber-laser detection, and developed the multi-step test procedure it runs. The measurement extracts Rayleigh-wave velocity to determine Young's modulus, and ties interparticle neck growth to elastic modulus through a grain-growth model. I validated it against impulse-excitation dynamic-resonance testing per ASTM.
The method is built for in-situ measurement. It has been demonstrated on room-temperature and post-process ceramic samples.
No manual existed for the system. I taught myself oscilloscope use and the twenty-five-step testing procedure to bring it to usable data, and processed the signals in Python and MATLAB with first-principles error analysis.
I operated a custom selective laser flash sintering platform for additive manufacturing of ceramics, developing scan recipes and optimizing the process parameters: laser power, scan speed, jump speed, and applied field voltage. Instrument control was written in ScanMaster ScanScript. Profilometry of the scan tracks identified cracking and set the usable laser-power window.
Upstream of that, I processed alumina and 8YSZ end to end. Powder processing, dry pressing, drying, and high-temperature sintering to 1650 °C, with sintering schedules designed to target controlled densification and validated by measurement.
I designed and fabricated the custom hardware the experiments needed, in machined aluminum, ceramic, and Teflon, including kinematic couplings for transferring samples between systems. I rebuilt and realigned the optical paths of both custom laser systems across four laser replacements, and redesigned and rewired a high-voltage system after diagnosing that the original wiring could not carry the operating voltage. I wrote the standard operating procedures for the Class IV laser system and specified and installed its controls.
For characterization I operated the SEM from load to unload on non-conductive ceramics, including the charge-mitigation preparation that it requires, along with optical profilometry and some quantitative image analysis.
I served as Lab Safety Lead, trained all new laboratory personnel, and mentored undergraduate researchers.